  • 4200.jpg

Total Organic Carbon Analyzer TCN-4200


Total organic matter determination in environmental water quality can be applied to tap water, surface water, rivers, lakes, seawater, domestic sewage, and industrial wastewater; organic content determination in new-energy battery materials, chemical products, and electroplating solutions ensures product quality and quality control; the ultraviolet oxidation method is suitable for detecting low-concentration organic matter in water samples, such as drinking water, boiler water, and purified water used in laboratory research; additionally, solid-phase and total nitrogen–total sulfur modules can be added to measure the content of organic matter, total nitrogen, and total sulfur in solid or liquid samples.


Product Details


Application


Total organic matter determination in environmental water quality can be applied to tap water, surface water, rivers, lakes, seawater, domestic sewage, and industrial wastewater; organic content analysis of new-energy battery materials, chemical products, and electroplating solutions ensures product quality and quality control; the ultraviolet oxidation method is suitable for detecting low-concentration organic matter in water samples, such as drinking water, boiler water, and purified water used in laboratory research; additionally, solid-phase and total nitrogen–total sulfur modules can be added to measure the content of organic matter, total nitrogen, and total sulfur in solid or liquid samples.

 

Product Features


  1. An eight-way valve can be used to automatically perform acid addition, sample injection, and dilution.
  2. The analyzer can employ either persulfate oxidation or high-temperature catalytic oxidation to determine total carbon, complying with a broader range of industry-standard methods.
  3. High-temperature catalytic furnace, which can be equipped with different catalysts depending on the sample; temperature can be set within a range of 680–900°C.
  4. A heated and thermostatically controlled inorganic carbon reaction unit eliminates peak tailing, shortens sample analysis time, and ensures the reproducibility of measurement results.
  5. Sample Injection Pump: Uses a syringe pump for sample injection, offering high precision and excellent reproducibility; the injection volume can be controlled according to the sample measurement list.
  6. Efficient Nafion tubing is used to effectively isolate water vapor, eliminating the need for conventional dehydration membranes and providing more effective prevention of gas-path blockage.
  7. Optional 24-position autosampler available for unattended operation, saving labor and time.
  8. Optional solid-sample introduction module for measuring solid and semi-solid samples.
  9. The detector’s gas chamber features an electroplated, high-infrared-reflectivity metal coating on its inner walls, minimizing infrared light loss and ensuring high detection accuracy.
  10. Front-mounted burner design for convenient operation and maintenance.
  11. The detector employs 24-bit A/D conversion technology, resulting in more stable data, lower noise, a wider detection range, and faster response.
  12. Long-tube, plug-and-play deionizer that effectively eliminates the impact of ions on the detector, eliminating the need for manual loading of the deionizing agent.
  13. Scalable total nitrogen detector for total nitrogen measurement
  14. Complies with the international standard ISO 8245, the National Environmental Protection Standard of the People’s Republic of China HJ 501-2009, and the National Metrological Verification Procedure of the People’s Republic of China JJG 821-2005.

Technical Specifications


Analysis modes: TC, TIC, NPOC, TOC

Degradation Principle TC: High-Temperature Catalytic Oxidation, UV-Persulfate Oxidation

TIC: 10% phosphoric acidification and aeration

Carrier gas: High-purity O2, 99.999%

Combustion temperature: adjustable between 680–1000°C

IC digestion temperature: 40℃

Catalyst: CeO2 or Pt/Al2O3

Measurement principle: NDIR